MX/MF
Test with the Best - True 0.01% (S.D.)
Products
A&D’s MX/MF-50 moisture analyzer boasts a major advantage over other moisture analyzer devices, namely, the ability to conduct analysis of a sample containing very little moisture content. This is advantageous when measuring samples with low evaporation characteristics. Side-by-side comparisons with the Karl Fisher method reveals the MX/MF-50 gives comparable results.
*MX-50 displays 0.01% resolution for moisture contents.
Moisture measurement result below is mean value of 5 measurements.
Sample : Plastic Polyethylene Terephthalate Pellet(PET)
Measurement condition
Heating Temp. : 180°C
Analyzers : Heating and drying method moisture analyzer : MX-50
Result
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
10 |
0.298 |
6.8 |
Standard-HI |
Sample : Lipstick
Measurement condition
Heating Temp. : 100°C
Analyzers : Heating and drying method moisture analyzer : MX-50
Result
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
1 |
0.778 |
1.9 |
Standard-MID |
Sample : Silica Sand
Measurement condition
Heating Temp. : 200°C
Analyzers : Heating and drying method moisture analyzer : MX-50
Result
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
10 |
0.498 |
2.3 |
Standard-HI |
Sample : Cement (Powder)
Measurement condition
Heating Temp. : 200°C
Analyzers : Heating and drying method moisture analyzer : MX-50
Result
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.408 |
3 |
Standard-MID |
Sample : Suger Crystals (Powder)
Measurement condition
Heating Temp. : 160°C
Analyzers : Heating and drying method moisture analyzer : MX-50
Result
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.162 |
1.7 |
Standard-MID |
Sample : Seasoned Salt
Measurement condition
Heating Temp. : 100°C
Analyzers : Heating and drying method moisture analyzer : MX-50
Result
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.086 |
1.1 |
Standard-MID |
Sample : Anhydrous Glucose
Measurement condition
Heating Temp. : 140°C
Analyzers : Heating and drying method moisture analyzer : MX-50
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.696 |
1.7 |
Standard-MID |
Sample : Powdered Cold Beverage
Measurement condition
Heating Temp. : 120°C
Analyzers : Heating and drying method moisture analyzer : MX-50
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.408 |
2.7 |
Standard-MID |
Sample : Zinc Oxide
Measurement condition
Heating Temp. : 200°C
Analyzers : Heating and drying method moisture analyzer : MX-50
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.142 |
2.3 |
Standard-HI |
Sample : Aluminum Oxide
Measurement condition
Heating Temp. : 200°C
Analyzers : Heating and drying method moisture analyzer : MX-50
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.098 |
2.4 |
Standard-HI |
Sample : Talc
Measurement condition
Heating Temp. : 200°C
Analyzers : Heating and drying method moisture analyzer : MX-50
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.144 |
2.5 |
Standard-HI |
Sample : Calcium Carbonate
Measurement condition
Heating Temp. : 200°C
Analyzers : Heating and drying method moisture analyzer : MX-50
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
5 |
0.144 |
2.5 |
Standard-HI |
Sample : Plastic ABS Resin
Measurement condition
Heating Temp. : 140°C
Analyzers : Heating and drying method moisture analyzer : MX-50
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
10 |
0.425 |
12.1 |
Automatic(0.05%/min) |
Sample : Plastic polymethyl methacrylate
Measurement condition
Heating Temp. : 100°C
Analyzers : Heating and drying method moisture analyzer : MX-50
|
Sample Weight(g) |
Moisture(%) |
Heating Time(min) |
Measurement Mode |
|
10 |
0.488 |
19.4 |
Automatic(0.05%/min) |